PICMUS is part of the IEEE IUS 2016, Tours, France.
One of the guideline of the simulation was to generate synthetic phantoms with close properties as the ones from real acquisitions. To this end, the following points have been taking into account:
Pitch | 0.30 mm |
Element width | 0.27 mm |
Element height | 5 mm |
Elevation focus | 20 mm |
Number of elements | 128 |
Aperture | 38.4 mm |
Transmit frequency | 5.208 MHz |
Sampling frequency | 20.832 MHz |
Pulse bandwidth | 67% |
Excitation | 2.5 cycles |
To be noted: because of the combined use of the elevation focus and an attenuation coefficient, a small delay of 0.2 mm along depth can be observed on the simulated phantom used to evaluate the performance on resolution and distorsion. Since this effect comes from the transmission scheme, it should not be compensated by the reconstructed algorithms.